Feasibility of thin-slice abdominal CT in overweight patients using a vendor neutral image-based denoising algorithm: Assessment of image noise, contrast, and quality


Autoři: Akio Tamura aff001;  Manabu Nakayama aff001;  Yoshitaka Ota aff002;  Masayoshi Kamata aff002;  Yasuyuki Hirota aff002;  Misato Sone aff001;  Makoto Hamano aff001;  Ryoichi Tanaka aff003;  Kunihiro Yoshioka aff001
Působiště autorů: Department of Radiology, Iwate Medical University School of Medicine, Morioka, Japan aff001;  Division of Central Radiology, Iwate Medical University Hospital, Morioka, Japan aff002;  Division of Dental Radiology, Department of General Dentistry, Iwate Medical University School of Dentistry, Morioka, Japan aff003
Vyšlo v časopise: PLoS ONE 14(12)
Kategorie: Research Article
doi: https://doi.org/10.1371/journal.pone.0226521

Souhrn

The purpose of this study was to investigate whether the novel image-based noise reduction software (NRS) improves image quality, and to assess the feasibility of using this software in combination with hybrid iterative reconstruction (IR) in image quality on thin-slice abdominal CT. In this retrospective study, 54 patients who underwent dynamic liver CT between April and July 2017 and had a body mass index higher than 25 kg/m2 were included. Three image sets of each patient were reconstructed as follows: hybrid IR images with 1-mm slice thickness (group A), hybrid IR images with 5-mm slice thickness (group B), and hybrid IR images with 1-mm slice thickness denoised using NRS (group C). The mean image noise and contrast-to-noise ratio relative to the muscle of the aorta and liver were assessed. Subjective image quality was evaluated by two radiologists for sharpness, noise, contrast, and overall quality using 5-point scales. The mean image noise was significantly lower in group C than in group A (p < 0.01), but no significant difference was observed between groups B and C. The contrast-to-noise ratio was significantly higher in group C than in group A (p < 0.01 and p = 0.01, respectively). Subjective image quality was also significantly higher in group C than in group A (p < 0.01), in terms of noise and overall quality, but not in terms of sharpness and contrast (p = 0.65 and 0.07, respectively). The contrast of images in group C was greater than that in group A, but this difference was not significant. Compared with hybrid IR alone, the novel NRS combined with a hybrid IR could result in significant noise reduction without sacrificing image quality on CT. This combined approach will likely be particularly useful for thin-slice abdominal CT examinations of overweight patients.

Klíčová slova:

Abdominal muscles – Aorta – Computed axial tomography – Imaging techniques – Liver and spleen scan – Noise reduction – Vendors – Infrared radiation


Zdroje

1. Mahesh M. MDCT physics: the basics—Technology, image quality and radiation dose. Philadelphia: Lippincott Williams & Williams; 2012, p. 2009

2. National Comprehensive Cancer Network. Clinical practice guidelines in oncology: hepatobiliary cancer. https://www.nccn.org/professionals/physician_gls/pdf/hepatobiliary.pdf

3. National Comprehensive Cancer Network. Clinical practice guidelines in oncology: pancreatic adenocarcinoma. https://www.nccn.org/professionals/physician_gls/pdf/pancreatic.pdf

4. Zins M, Matos C, Cassinotto C. Pancreatic Adenocarcinoma Staging in the Era of Preoperative Chemotherapy and Radiation Therapy. Radiology. 2018;287 : 374–390. doi: 10.1148/radiol.2018171670 29668413

5. Sprawls P. AAPM tutorial. CT image detail and noise. Radiographics. 1992;12 : 1041–6. doi: 10.1148/radiographics.12.5.1529128 1529128

6. Schindera ST, Torrente JC, Ruder TD, Hoppe H, Marin D, Nelson RC, et al. Decreased detection of hypovascular liver tumors with MDCT in obese patients: a phantom study. AJR Am J Roentgenol. 2011;196: W772–W776. doi: 10.2214/AJR.10.5351 21606267

7. Funama Y, Awai K, Miyazaki O, Nakayama Y, Goto T, Omi Y, et al. Improvement of low-contrast detectability in low-dose hepatic multidetector computed tomography using a novel adaptive filter: evaluation with a computer-simulated liver including tumors. Invest Radiol. 2006;41 : 1–7. doi: 10.1097/01.rli.0000188026.20172.5d 16355033

8. Christianson O, Winslow J, Frush DP, Samei E. Automated Technique to Measure Noise in Clinical CT Examinations. AJR Am J Roentgenol. 2015;205:W93–9. doi: 10.2214/AJR.14.13613 26102424

9. Volders D, Bols A, Haspeslagh M, Coenegrachts K. Model-based iterative reconstruction and adaptive statistical iterative reconstruction techniques in abdominal CT: comparison of image quality in the detection of colorectal liver metastases. Radiology. 2013;269 : 469–474. doi: 10.1148/radiol.13130002 23847252

10. Smith EA, Dillman JR, Goodsitt MM, Christodoulou EG, Keshavarzi N, Strouse PJ. Model-based iterative reconstruction: effect on patient radiation dose and image quality in pediatric body CT. Radiology. 2014;270 : 526–534. doi: 10.1148/radiol.13130362 24091359

11. Kataria B, Althén JN, Smedby Ö, Persson A, Sökjer H, Sandborg M. Assessment of image quality in abdominal CT: potential dose reduction with model-based iterative reconstruction. Eur Radiol. 2018;28 : 2464–2473. doi: 10.1007/s00330-017-5113-4 29368163

12. Minamishima K, Sugisawa K, Yamada Y, Jinzaki M. Quantitative and qualitative evaluation of hybrid iterative reconstruction, with and without noise power spectrum models: A phantom study. J Appl Clin Med Phys. 2018;19 : 318–325. doi: 10.1002/acm2.12304 29493077

13. Brennan DD, Zamboni GA, Raptopoulos VD, Kruskal JB. Comprehensive preoperative assessment of pancreatic adenocarcinoma with 64-section volumetric CT. Radiographics. 2007;27 : 1653–1666. doi: 10.1148/rg.276075034 18025509

14. Oguro S, Funabiki T, Hosoda K, et al. 64-Slice multidetector computed tomography evaluation of gastrointestinal tract perforation site: detectability of direct findings in upper and lower GI tract. Eur Radiol. 2010;20 : 1396–403. doi: 10.1007/s00330-009-1670-5 19997849

15. Abdelmoumene A, Chevallier P, Chalaron M, Schneider F, Verdun FR, Frascarolo P, et al. Detection of liver metastases under 2 cm: comparison of different acquisition protocols in four row multidetector-CT (MDCT). Eur Radiol. 2005;15 : 1881–1887. doi: 10.1007/s00330-005-2741-x 15868125

16. Soo G, Lau KK, Yik T, Kutschera P. Optimal reconstructed section thickness for the detection of liver lesions with multidetector CT. Clin Radiol. 2010;65 : 193–197. doi: 10.1016/j.crad.2009.10.009 20152274

17. Boedeker KL, McNitt-Gray MF. Application of the noise power spectrum in modern diagnostic MDCT: part II. Noise power spectra and signal to noise. Phys Med Biol. 2007;52 : 4047–4061. doi: 10.1088/0031-9155/52/14/003 17664594

18. Bushberg JT, Seibert JA, Leidholdt EM, et al. The essential physics of medical imaging. 2nd ed. Philadelphia, Lippincott Williams & Wilkins; 2002, PA:369

19. Sagara Y, Hara AK, Pavlicek W, Silva AC, Paden RG, Wu Q. Abdominal CT: comparison of low-dose CT with adaptive statistical iterative reconstruction and routine-dose CT with filtered back projection in 53 patients. AJR Am J Roentgenol. 2010;195 : 713–719. doi: 10.2214/AJR.09.2989 20729451

20. Hara AK, Wellnitz CV, Paden RG, Pavlicek W, Sahani DV. Reducing body CT radiation dose: beyond just changing the numbers. AJR Am J Roentgenol. 2013;201 : 33–40. doi: 10.2214/AJR.13.10556 23789656

21. Schindera ST, Odedra D, Raza SA, Kim TK, Jang HJ, Szucs-Farkas Z, et al. Iterative reconstruction algorithm for CT: can radiation dose be decreased while low-contrast detectability is preserved? Radiology. 2013;269 : 511–518. doi: 10.1148/radiol.13122349 23788715

22. Schindera ST, Odedra D, Mercer D, Thipphavong S, Chou P, Szucs-Farkas Z, et al. Hybrid iterative reconstruction technique for abdominal CT protocols in obese patients: assessment of image quality, radiation dose, and low-contrast detectability in a phantom. AJR Am J Roentgenol. 2014;202: W146–W152. doi: 10.2214/AJR.12.10513 24450696

23. Del Gaizo AJ, Fletcher JG, Yu L, Paden RG, Spencer GC, Leng S, et al. Reducing radiation dose in CT enterography. Radiographics. 2013;33 : 1109–1124. doi: 10.1148/rg.334125074 23842974

24. Hara AK, Paden RG, Silva AC, et al. Iterative reconstruction technique for reducing body radiation dose at CT: feasibility study. AJR 2009; 193 : 764–771. doi: 10.2214/AJR.09.2397 19696291

25. Manduca A, Yu L, Trzasko JD, et al. Projection space denoising with bilateral filtering and CT noise modeling for dose reduction in CT. Med Phys 2009; 36 : 4911–4919. doi: 10.1118/1.3232004 19994500

26. Thibault JB, Sauer KD, Bouman CA, Hsieh J. A three-dimensional statistical approach to improved image quality for multislice helical CT. Med Phys 2007; 34 : 4526–4544 doi: 10.1118/1.2789499 18072519

27. Singh S, Kalra MK, Hsieh J, Licato PE, Do S, Pien HH, et al. Abdominal CT: comparison of adaptive statistical iterative and filtered back projection reconstruction techniques. Radiology. 2010;257 : 373–383. doi: 10.1148/radiol.10092212 20829535

28. Ehman EC, Guimarães LS, Fidler JL, et al. Noise reduction to decrease radiation dose and improve conspicuity of hepatic lesions at contrast-enhanced 80-kV hepatic CT using projection space denoising. AJR Am J Roentgenol. 2012;198 : 405–11. doi: 10.2214/AJR.11.6987 22268185

29. Marin D, Nelson RC, Schindera ST, Richard S, Youngblood RS, Yoshizumi TT, et al. Low-tube-voltage, high-tube-current multidetector abdominal CT: improved image quality and decreased radiation dose with adaptive statistical iterative reconstruction algorithm—initial clinical experience. Radiology. 2010;254 : 145–153. doi: 10.1148/radiol.09090094 20032149

30. Nakaura T, Awai K, Oda S, et al. Low-kilovoltage, high-tube-current MDCT of liver in thin adults: pilot study evaluating radiation dose, image quality, and display settings. AJR Am J Roentgenol. 2011;196 : 1332–8. doi: 10.2214/AJR.10.5698 21606297


Článek vyšel v časopise

PLOS One


2019 Číslo 12
Nejčtenější tento týden
Nejčtenější v tomto čísle
Kurzy

Zvyšte si kvalifikaci online z pohodlí domova

BONE ACADEMY 2025
nový kurz
Autoři: prof. MUDr. Pavel Horák, CSc., doc. MUDr. Ludmila Brunerová, Ph.D, doc. MUDr. Václav Vyskočil, Ph.D., prim. MUDr. Richard Pikner, Ph.D., MUDr. Olga Růžičková, MUDr. Jan Rosa, prof. MUDr. Vladimír Palička, CSc., Dr.h.c.

Cesta pacienta nejen s SMA do nervosvalového centra
Autoři: MUDr. Jana Junkerová, MUDr. Lenka Juříková

Svět praktické medicíny 2/2025 (znalostní test z časopisu)

Eozinofilní zánět a remodelace
Autoři: MUDr. Lucie Heribanová

Hypertrofická kardiomyopatie: Moderní přístupy v diagnostice a léčbě
Autoři: doc. MUDr. David Zemánek, Ph.D., MUDr. Anna Chaloupka, Ph.D.

Všechny kurzy
Kurzy Podcasty Doporučená témata Časopisy
Přihlášení
Zapomenuté heslo

Zadejte e-mailovou adresu, se kterou jste vytvářel(a) účet, budou Vám na ni zaslány informace k nastavení nového hesla.

Přihlášení

Nemáte účet?  Registrujte se

#ADS_BOTTOM_SCRIPTS#