Clinical Results Following the Implantation of a New Toric Intraocular Lens with Extended Depth of Focus
Clinical Results Following the Implantation of a New Toric Intraocular Lens with Extended Depth of Focus
Purpose: To evaluate the clinical outcomes following the implantation of a new toric intraocular lens (TIOL) with extended depth of focus (EDOF) Bi-Flex ELON Toric POB-MA 877PETY (Elon®, Medicontur Medical Engineering Ltd., Zsámbék, Hungary), including analysis of lens rotational stability and assessment of patient satisfaction.
Material and methods: In a retrospective, monocentric clinical study, 29 eyes of 18 patients (11 men and 7 women) with presbyopia, cataracts, and corneal astigmatism were evaluated. These patients underwent uncomplicated cataract surgery with Bi-Flex ELON Toric TIOL implantation. Postoperative assessments included measurements of monocular uncorrected and corrected distance visual acuity (UDVA, CDVA) and near visual acuity (UNVA, CNVA), rotational stability of the TIOL, and subjective evaluation of visual quality using a questionnaire. Follow-up examinations were conducted 6 months postoperatively (ranging from 3 to 9 months).
Results: The mean postoperative monocular UDVA improved from 0.4 ±0.2 to 0.90 ±0.1 (p < 0.0001). Average CDVA was 1.0 ±0.2. All patients achieved monocular UDVA ≥ 0.6 and CDVA ≥ 0.8. The average postoperative monocular UNVA was at Jaeger chart level 4, and CNVA reached Jaeger level 1 in all patients. The mean residual astigmatic refraction was -0.6 ±0.4 Dcyl (p = 0.0003), and the mean absolute rotational deviation was 1.9 ±2.65°. Patient satisfaction was assessed on a scale of 1 to 5, with an average score of 1.4. Targeted questioning revealed that halo-type dysphotopsia was reported in 52 % of eyes, glare-type in 17%, and no dysphotopsia in 31%.
Conclusion: Implantation of a Bi-Flex ELON Toric TIOL for correcting astigmatism and presbyopia significantly improved both distance and near visual acuity. The rotational stability of the lens was high, and the majority of patients achieved independence from wearing spectacles.
Keywords:
cataract – toric intraocular lens – extended depth of focus – IOL rotation
Autoři:
Simona Jančárová 1; Marie Němcová 1; Pavel Rozsíval 1,2; Viktor Ladislav Nováček 1
Působiště autorů:
Department of Ophthalmology, Institute of Aviation Medicine, Prague, Czech Republic
1; Charles University, Faculty of Medicine in Hradec Králové, Department of, Ophthalmology, University Hospital Hradec Králové, Czech Republic
2
Vyšlo v časopise:
Čes. a slov. Oftal., 81, 2025, No. Ahead of Print, p. 1-6
Kategorie:
Původní práce
doi:
https://doi.org/10.31348/2025/25
Souhrn
Purpose: To evaluate the clinical outcomes following the implantation of a new toric intraocular lens (TIOL) with extended depth of focus (EDOF) Bi-Flex ELON Toric POB-MA 877PETY (Elon®, Medicontur Medical Engineering Ltd., Zsámbék, Hungary), including analysis of lens rotational stability and assessment of patient satisfaction.
Material and methods: In a retrospective, monocentric clinical study, 29 eyes of 18 patients (11 men and 7 women) with presbyopia, cataracts, and corneal astigmatism were evaluated. These patients underwent uncomplicated cataract surgery with Bi-Flex ELON Toric TIOL implantation. Postoperative assessments included measurements of monocular uncorrected and corrected distance visual acuity (UDVA, CDVA) and near visual acuity (UNVA, CNVA), rotational stability of the TIOL, and subjective evaluation of visual quality using a questionnaire. Follow-up examinations were conducted 6 months postoperatively (ranging from 3 to 9 months).
Results: The mean postoperative monocular UDVA improved from 0.4 ±0.2 to 0.90 ±0.1 (p < 0.0001). Average CDVA was 1.0 ±0.2. All patients achieved monocular UDVA ≥ 0.6 and CDVA ≥ 0.8. The average postoperative monocular UNVA was at Jaeger chart level 4, and CNVA reached Jaeger level 1 in all patients. The mean residual astigmatic refraction was -0.6 ±0.4 Dcyl (p = 0.0003), and the mean absolute rotational deviation was 1.9 ±2.65°. Patient satisfaction was assessed on a scale of 1 to 5, with an average score of 1.4. Targeted questioning revealed that halo-type dysphotopsia was reported in 52 % of eyes, glare-type in 17%, and no dysphotopsia in 31%.
Conclusion: Implantation of a Bi-Flex ELON Toric TIOL for correcting astigmatism and presbyopia significantly improved both distance and near visual acuity. The rotational stability of the lens was high, and the majority of patients achieved independence from wearing spectacles.
Štítky
OftalmologieČlánek vyšel v časopise
Česká a slovenská oftalmologie

2025 Číslo Ahead of Print
- Diagnostický algoritmus při podezření na syndrom periodické horečky
- Stillova choroba: vzácné a závažné systémové onemocnění
- Léčba zánětů spojivek a mazových žlázek víčka v primární péči
- Familiární středomořská horečka
- Citikolin jako užitečný pomocník v léčbě diabetické retinopatie a glaukomu
Nejčtenější v tomto čísle
- Idiopathic Corneal Thinning and Spontaneous Perforation at Early Adulthood. A Case Report
- Uveitis associated with juvenile idiopathic arthritis (JIA) and JIA-like uveitis in pediatric patients
- Rhegmatogenous Retinal Detachment in Childhood
- Reconstruction of the Anophthalmic Conjunctival Sac. A Review of Surgical Procedures to Achieve Stability of the Ocular Prosthesis in Our Practice