Verification of the reference range of free triiodothyronine in serum on the analyzer Architect ci 16200


Authors: T. Vacková;  J. Franeková;  A. Jabor
Authors‘ workplace: Praha ;  Oddělení klinické biochemie, Pracoviště laboratorních metod, Institut klinické a experimentální medicíny
Published in: Klin. Biochem. Metab., 26, 2018, No. 4, p. 178-181

Overview

Objective:

Verification of the reference interval of a new method for the determination of free triiodothyronine in serum on the Abbott Architect 16200 analyzer.

Design:

Original Article.

Settings:

Institute for Clinical and Experimental Medicine, Vídeňská 1958/9, 140 21 Praha

Material and methods:

The reference interval was validated on a total of 110 serum samples from healthy subjects, comprising 55 females and 55 males. All samples used were from the IKEM Biobanka. The FT3_6 microparticle chemiluminescent immunoassay Abbott kit was used for the assay.

Results:

All free serum triiodothyronine values measured by the new fT3_6 method were in the reference interval indicated by the manufacturer. The reference range determined by a non-parametric statistical method with the Box-Cox transformation on the IKEM Biobanka samples was 3.21 -⁠ 4.78 pmol/L.

Conclusion:

The fT3 reference range established by the Abbott’s new fT3_6 method can be used in clinical practice with the range specified by the manufacturer.

Keywords:

Architect 16200, fT3, reference interval (RI). 


Sources

1. Abbott Diagnostic Division Informace o produktu ARCHITECT Free T3, Ireland, 13. 7. 2017, 11 s.

2. Abbott Diagnostic Division Příbalový leták k produktu ARCHITECT Free T3 platný pro katalogová čísla 7K63-27(37, 32), česká verze G8-1469/R03, 03/2017, 6 s.

3. Límanová, Z., Pikner, R., Springer, D. Doporučení pro laboratorní diagnostiku funkčních a autoimunitních onemocnění štítné žlázy. Klin. Biochem. Metab., 19 (40), 2011, No. 1, p. 48-61.

4. Demers, L.M., Spencer, C.A. Laboratory Support for the Diagnosis and Monitoring of Thyroid Disease 2002 NACB, Laboratory Medicine Practise Guidelines, 2002, 13, p. 125

5. Zima, T. Laboratorní diagnostika. Praha: Galén, 2013, 1146 s., ISBN 9788074920622

6. Thomas, L. Clinical Laboratory Diagnostics: Use and Assessment of Clinical Laboratory Results. Frankfurt am Main: TH-Books, 1998, 1527 s. ISBN 3-9805215-4-0.

7. Röcker, L., Janiszewski, Ch., Hopfenmüller, W. Determination of reference ranges for thyroid hormones in serum for the population of Berlin (West) with special emphasis of age -⁠ and gender -⁠ dependence. Lab. Med. 1988, 12, p. 213-22

8. Liappis, N., Bernstorff v, W., Schlebusch, H. Refe-rence values for the concentration of free triiodothyronine, triiodothyronine, free thyroxine, thyroxine, thyrotropin and thyroxin-binding globulin in serum of euthyroid adults. Influence of oral contraceptives. Method: luminiscence-enhanced enzyme immunoassay. Ärztl. Lab. 1990, 36, p. 8-11

9. Racek, J. Klinická biochemie. Praha: Galén, 2006, 329 s. ISBN 8072623249

10. Abbott Diagnostic Division Příbalový leták k produktu ARCHITECT Free T4 platný pro katalogová čísla 7K65-39 (24, 39, 34), česká verze G8-1500/R03, 04/2017, 6 s.

11. Abbott Diagnostic Division Příbalový leták k produktu ARCHITECT TSH platný pro katalogová čísla 7K62-35 (20, 25, 30), česká verze G6-0153/R06, 11/2015, 7 s.

Labels
Clinical biochemistry Nuclear medicine Nutritive therapist
Login
Forgotten password

Enter the email address that you registered with. We will send you instructions on how to set a new password.

Login

Don‘t have an account?  Create new account

#ADS_BOTTOM_SCRIPTS#