Retrospective analysis of the results of an external quality assessment of the determination of bilirubin in newborn serum using a unified calibrator.


Authors: Š. Hajzer
Published in: Klin. Biochem. Metab., 34, 2026, No. 3, p. 0

Overview

Objective: Retrospective analysis of the impact of a unified calibrator on the interlaboratory variability of the results of the total bilirubin determinations in newborn serum.

Design: retrospective cross-sectional
Material and Methods: Retrospective reassessment of the results of participants analyses´ in the external quality assessment (EHK) in the period 2023 to 2025 using the unified calibrator and graphical and statistical evaluation of the results.

Results: Reassessment of the three-year results of the EHK system provided by SEKK s.r.o. and their statistical analysis demonstrated a significant positive impact of implementing a single calibrator on reducing interlaboratory variability and the degree of harmonization of analytical platforms. The data analysis focused on the four dominant platforms of Abbott, Beckmann, Roche and Siemens, on which about 95% of all neonatal bilirubin determinations are performed in clinical labo -⁠ ratories. The original three-year pooled datasets exhibited significant variability in results, with relative bias values ranging up to ±22%. The most pronounced differences in trueness were identified between the Roche (Cobas) and Siemens (ADVIA and Atellica) platforms, with median relative biases of -5.45% (interquartile range, IQR 5.4) and +12.6% (IQR 10.35), respectively. However, the application of a single calibrator led to a significant statistical correction by reducing the variance for all four platforms. The average interquartile range (IQR) decreased on average to 4.1, indicating an increase in precision, and the median relative bias stabilized within a narrow interval around zero, thus stabilizing both the trueness and the overall accuracy of the analyses. Retrospective analysis of biases demonstrated long-term systematic differences between the diagnostics of dominant manufacturers. The main source of undesirable variability in the determination of bilirubin in newborn serum is not the chemical principle of the method itself, but inconsistent calibration of IVD products.

Conclusion: The results of our paper show that the implementation of a unified calibration system is of key importance for the standardization of measurements of total bilirubin in newborn serum. However, there is still a need to define a unified analytical procedure, metrological traceability, redefinition of adequate permissible limits of Dmax, special calibrators and control materials suitable for the purpose of measurement.

Keywords:

bilirubin – standardization – hyperbilirubinemia – reference method


Sources

1. Thomas, D. H., Warner, J. V., Jones, G. R.D., et al. Total bilirubin assay differences may cause inconsistent treatment decisions in neonatal hyperbilirubinaemia. J. Clin. Chem. Lab. Med., 2022, 60(11), 1736-1744. odkaz

2. Oostendorp, M., ten Hove, C. H., van Berkel, M., Roovers, L. A Significant Increase in the Incidence of Neonatal Hyperbilirubinemia and Phototherapy Treatment Due to a Routine Change in Laboratory Equipment. Arch. Pathol. Lab. Med., 2024, 148, e40-e47 odkaz

3. Hulzebos, C. V., Camara, J. E., van Berkel, M. et al. and (IFCC WG-NB). Bilirubin measurements in neonates: uniform neonatal treatment can only be achieved by improved standardizataion. Clin. Chem. Lab. Med., 2024, 62(10), 1892-1903. odkaz

4. Hajzer Š. Súčasný stav stanovenie celkového bilirubínu v sére novorodencov. Klin. Biochem. Metab., 2024, 32(4), 90-96 odkaz

5. Westenberg, L. E.H., van Schrojenstein Lantman, M., van Berkel, M. et al. Evaluation of Total Bilirubin in a National Quality Program. Clin. Chem., 2026, 72(1), 183 -191. [odkaz]

6. Bhutani, V. K., Johnson, L., Sivieri, E. M. Predictive ability of a predischarge hour-specific serum bilirubin for subsequent significant hyperbilirubinemia ih healthy term and near-term newborns. Pediadtrics, 1999, 103(1), 6-14. odkaz

7. Kittanakom, S., Sea, J., Leung, F. et al. Impact of switching total bilirubin assays on the classification of neonates at high risk for hyperbilirubinemia. Clin. Chem. Lab. Med. (CCLM), 2023, 61(9), e175-e178. odkaz

8. Hynková, P., Kučerová, V., Pospíšilová, J. Srovnání meření celkového bilirubinu na analyzátoru Cobas 8000 versus Atellica CH v centrální laboratoři, versus měření na analyzátoru v režimu POCT. Klin. Biochem. Metab., 2023, 31(3), 118. odkaz

9. Tokuda; Kuniaki et al. Method for measuring bilirubin. United States Patent: US-5449623 -A Publication date 1995-09-12 odkaz

10. Záverečná zpráva k vyhodnocení cyklu EHK: BIL2/24: Bilirubin novorozenecký. SEKK, 2024. odkaz

11. Roehle, G., Schlebusch, H., Geilenkeuser, W. J., Kruse, R. External quality control in the determination of neonatal bilirubin. An approach to the improvement of results. J Clin. Chem. Clin. Biochem., 1988, 26(7), 441-446. odkaz

12. Hajzer, S., Osadská, E., Záchenská, A. Evaluation of interlaboratory proficiency surveys of bilirubin determinations in sera of newborns. Eur. J Clin. Chem. Clin. Biochem., 1992, 30(5), 291-295. odkaz

13. Friedecký, B. Harmonizace v klinické laboratoŕi. Pokus o pohled s nadhledem. Klin. Biochem. Metab., 2024, 32(4), 88-89. odkaz

Labels
Clinical biochemistry Nuclear medicine Nutritive therapist
Login
Forgotten password

Enter the email address that you registered with. We will send you instructions on how to set a new password.

Login

Don‘t have an account?  Create new account

#ADS_BOTTOM_SCRIPTS#