Intraocular pressure according to different types of tonometry (non-contact and Goldmann applanation) in patients with different degrees of bilateral tearing


Autoři: Bo Ram Seol aff001;  Tae Gu Kang aff001;  Bonhyeok Gu aff001
Působiště autorů: Department of Ophthalmology, Veterans Health Service (VHS) Medical Center, Seoul, Korea aff001
Vyšlo v časopise: PLoS ONE 14(9)
Kategorie: Research Article
doi: https://doi.org/10.1371/journal.pone.0222652

Souhrn

Purpose

To investigate intraocular pressure (IOP) readings by non-contact tonometry (NCT) and Goldmann applanation tonometry (GAT) for patients with different degrees of bilateral tearing.

Methods

In this study, we reviewed the medical charts of patients complaining of different degrees of bilateral tearing. The tear meniscus height (TMH) and IOP with NCT and GAT were measured. In each patient, a comparison of IOP readings between the eye with lower TMH and the contralateral eye with higher TMH was evaluated. The TMH was graded as follows: grade 1 (low): TMH < 0.2 mm; grade 2 (moderate): 0.2 mm ≤ TMH < 0.6 mm; grade 3 (high): TMH ≥ 0.6 mm. Subsequently, a comparison of IOP readings among eyes with low, moderate, and high TMH was also performed.

Results

A total of 120 eyes of 60 patients were enrolled. When comparing the two eyes of a patient, the eye with higher TMH showed higher NCT readings and larger difference in IOP readings between the two tonometries than the eye with lower TMH (P < 0.001 and P < 0.001, respectively). When TMH was classified into grades according to the degree, the high TMH eyes showed higher NCT readings than did the low and moderate TMH eyes (P < 0.001 and P = 0.001, respectively). In addition, the high TMH eyes showed a larger difference in IOP readings between the two tonometries than did the low and moderate TMH eyes (P < 0.001 and P < 0.001, respectively).

Conclusion

Eyes with higher TMH showed higher NCT readings and a larger difference in IOP between the two tonometries (NCT and GAT) than those with lower TMH. In patients with tearing, the NCT value may be inaccurate, so it is necessary to measure the GAT.

Klíčová slova:

Biology and life sciences – Anatomy – Head – Cornea – Intraocular pressure – Medicine and health sciences – Eyes – Ocular system – Ocular anatomy – Ophthalmology – Eye diseases – Glaucoma – Endocrinology – Endocrine disorders – Metabolic disorders – Vascular medicine – Blood pressure – Hypertension – Health care – Health care providers – Medical doctors – Physicians – Computer and information sciences – Data visualization – Infographics – Charts – People and places – Population groupings – Professions – Medical personnel


Zdroje

1. Yilmaz I, Altan C, Aygit ED, Alagoz C, Baz O, Ahmet S, et al. Comparison of three methods of tonometry in normal subjects: Goldmann applanation tonometer, non-contact airpuff tonometer, and Tono-Pen XL. Clin Ophthalmol. 2014;8:1069–74. doi: 10.2147/OPTH.S6391 24944507

2. Farhood QK. Comparative evaluation of intraocular pressure with an air-puff tonometer versus a Goldmann applanation tonometer. Clin Ophthalmol. 2013;7:23–7. doi: 10.2147/OPTH.S38418 23293511

3. Martinez-de-la-Casa JM, Jimenez-Santos M, Saenz-Frances F, Matilla-Rodero M, Mendez-Hernandez C, Herrero-Vanrell R, et al. Performance of the rebound, noncontact and Goldmann applanation tonometers in routine clinical practice. Acta Ophthalmol. 2011;89(7):676–80. doi: 10.1111/j.1755-3768.2009.01774.x 19900196.

4. Rand Allingham R KF, Sharon F, et al. Shield's textbook of glaucoma. 6th ed. Philadelphia; 2005.

5. Grolman B. A new tonometer system. Am J Optom Arch Am Acad Optom.1972;49(8):646–60. 4506671.

6. Sagan W, Schwaderer K. Non-contact tonometry by assistants. Am J Optom Physiol Opt. 1975;52(4):288–90. 236657.

7. Cook JA, Botello AP, Elders A, Fathi Ali A, Azuara-Blanco A, Fraser C, et al. Systematic review of the agreement of tonometers with Goldmann applanation tonometry. Ophthalmology. 2012;119(8):1552–7. doi: 10.1016/j.ophtha.2012.02.030 22578443.

8. Ehlers N, Bramsen T, Sperling S. Applanation tonometry and central corneal thickness. Acta Ophthalmol. 1975;53(1):34–43. doi: 10.1111/j.1755-3768.1975.tb01135.x 1172910.

9. Ehlers N, Hansen FK, Aasved H. Biometric correlations of corneal thickness. Acta Ophthalmol. 1975;53(4):652–9. doi: 10.1111/j.1755-3768.1975.tb01784.x 1242286.

10. Whitacre MM, Stein R. Sources of error with use of Goldmann-type tonometers. Surv Ophthalmol. 1993;38(1):1–30. 8235993.

11. Damji KF, Muni RH, Munger RM. Influence of corneal variables on accuracy of intraocular pressure measurement. J Glaucoma. 2003;12(1):69–80. 12567117.

12. Tonnu PA, Ho T, Newson T, El Sheikh A, Sharma K, White E, et al. The influence of central corneal thickness and age on intraocular pressure measured by pneumotonometry, non-contact tonometry, the Tono-Pen XL, and Goldmann applanation tonometry. Br J Ophthalmol. 2005;89(7):851–4. doi: 10.1136/bjo.2004.056622 15965165.

13. Ito K, Tawara A, Kubota T, Harada Y. IOP measured by dynamic contour tonometry correlates with IOP measured by Goldmann applanation tonometry and non-contact tonometry in Japanese individuals. J Glaucoma. 2012;21(1):35–40. doi: 10.1097/IJG.0b013e31820275b4 21173706.

14. Jones LT. The lacrimal secretory system and its treatment. Am J Ophthalmol. 1966;62(1):47–60. doi: 10.1016/0002-9394(66)91676-x 5936526.

15. Allingham RR, Shields MB. Shields’ textbook of glaucoma. 6th. Wolters Kluwer/Lippincott Williams & Wilkins Health; Philadelphia: 2011.

16. Kawai M, Yamada M, Kawashima M, Inoue M, Goto E, Mashima Y, et al. Quantitative evaluation of tear meniscus height from fluorescein photographs. Cornea. 2007;26(4):403–6. doi: 10.1097/ICO.0b013e318033c242 17457186.

17. Burkat CN, Lucarelli MJ. Tear meniscus level as an indicator of nasolacrimal obstruction. Ophthalmology. 2005;112(2):344–8. doi: 10.1016/j.ophtha.2004.07.030 15691573.

18. Forbes M, Pico G Jr, Grolman B. A non-contact tonometer. Arch Ophthalmol. 1974;91(2):134–40. doi: 10.1001/archopht.1974.03900060140012 4810646.

19. Sorensen PN. The non-contact tonometer. Clinical evaluation on normal and diseased eyes. Acta Ophthalmol. 1975;53(4):513–21. doi: 10.1111/j.1755-3768.1975.tb01771.x 1242275


Článek vyšel v časopise

PLOS One


2019 Číslo 9
Nejčtenější tento týden
Nejčtenější v tomto čísle
Kurzy Podcasty Doporučená témata Časopisy
Přihlášení
Zapomenuté heslo

Zadejte e-mailovou adresu, se kterou jste vytvářel(a) účet, budou Vám na ni zaslány informace k nastavení nového hesla.

Přihlášení

Nemáte účet?  Registrujte se

#ADS_BOTTOM_SCRIPTS#